Installing the Agilent N4916B

4

Open the 

Help

 menu 

and click 

About

.

3

Check the Software Status of your Serial BERT N4903B

The software revision must be
7.0 or later. If it is lower, you 
need to update the software. 
For details refer to the User 
Guide.

www.agilent.com/find/n4903

Inspect the 

Agilent 

IO Control

 icon in the 

Windows task bar.

5

If you see this icon, 
the revision of the 
Agilent IO Libraries 
Suite is 15.5 or later.

You may wish to connect a keyboard and
mouse to your Serial BERT N4903B.
This is not mandatory but convenient
and should be done when the instrument
is switched off.

1

Open the 

Utility

 menu 

and click 

Minimize GUI

.

4

Switch on the Serial BERT 
N4903B.

2

You can use the 

Agilent

Connection Expert

 which

makes it easy to configure
USB instruments.

Содержание N4916B

Страница 1: ...nline Help of your High Performance Serial BERT N4903B For more information visit The Help also offers printable versions of the User Guide and the Programming Guide www agilent com find jbert You onl...

Страница 2: ...I pollution degree 2 environment It is designed to operate within a temperature range of 5 40 C 40 105 F at a maximum relative humidity of 95 and at altitudes of up to 2000 meters Refer to the specifi...

Страница 3: ...the installation media All data and clock input and output connectors of the PG that are not used in your test setup must be terminated with 50 Ohm The instrument can operate from any single phase AC...

Страница 4: ...ct the Agilent IO Control icon in the Windows task bar 5 If you see this icon the revision of the Agilent IO Libraries Suite is 15 5 or later You may wish to connect a keyboard and mouse to your Seria...

Страница 5: ...he rear USB ports of your N4903B 3 When you connect the N4916B for the first time the Windows Found New Hardware Wizard appears and asks whether it should connect to Windows Update Answer No 4 Click N...

Страница 6: ...e alias name to DATA_OUT or AUXDATA_OUT depending on the channel Data Out and Aux Data Out used This is the alias used by the software 3 After clicking OK the Agilent Connection Expert shows a window...

Страница 7: ...these functions 5 Enabling Clock Multiplier Select the Clock Multiplier function to open the Clock Multiplier Connection dialog 7 If you select the Clock Multiplier function a new menu entry is added...

Страница 8: ...For details see the Help and the User Guide Choose a suitable pattern e g PRBS and start the pattern generator 9 After some adjustments of the scope you will see a display like the following 10 In ca...

Страница 9: ...of problems please refer to the section Problems with the N4916B of the Help and the User Guide of your N4903B Preset the instrument state 1 Preset the instrument state 1 Enable Clock Multiplier from...

Отзывы: