Agilent Technologies N4916B Скачать руководство пользователя страница 7

Checking the Agilent N4916B

7

4

Disable the outputs of the 
J-BERT and use a 2.4 mm 
cable (m-m) and connect 
the pattern generator's 
Data Out port to the 
Input port of the N4916B.

1

Mount the SMA 50 Ohm 
termination on the unused 
Data Out port of the pattern 
generator using the SMA 
to 2.4 mm adapter.

2

Restore the N4903B user
interface from the 
Windows taskbar.

3

Connect the N4916B to the Pattern Generator

From the Navigation Menu, 
select 

External Instrument(s)

 

and then click 

Config

 to

access 

Config

 window.

The

 Config

 window displays a list of instruments

externally connected  to the Serial BERT N4903B
along with the functions associated with them.

If the setup is used for the first time, perform the 
calibrated settings. Refer to the help system for 
detailed calibration instructions.

Let’s follow the next steps to enable these functions.

5

Enabling Clock Multiplier

Select the 

Clock Multiplier

 function to open 

the 

Clock Multiplier Connection

 dialog.

7

If you select the Clock Multiplier function,
a new menu entry is added to the 

External

Instrument(s)

 submenu. 

From the Navigation Menu, select 

External Instrument(s)

 and then click

Clock Multiplier-DATAOUT

 to access 

Clock Multiplier-DATAOUT

 window. 

7

8

Set the values as following:
- Input Frequency: 1 to 7

 

.6 GHz

- Termination Voltage: -2.0 V to 2.0 V

5

Select the 

De-Emphasis

 function to open 

the 

De-Emphasis Connection

 dialog.

Enabling De-Emphasis

Press 

Enable

 to activate the connection between the N4903B 

and the N4916B and also enables the output of the J-BERT.

6

A selected checkbox corresponding to the
function indicates that the connection is 
established between the N4903B and 
N4916B.

From the 

Navigation Menu

, select 

PG Setup 

and click 

Data Output

 to access

Data Output

 window. 

8

Set the values as following:
- Pre-Cursor : 

 

- Post-Cursor1 : -
- Post-Cursor2 : 

0 dB to 12.04 dB

 

12.04 dB to 0 dB

 

-8 dB to 0 dB

Содержание N4916B

Страница 1: ...nline Help of your High Performance Serial BERT N4903B For more information visit The Help also offers printable versions of the User Guide and the Programming Guide www agilent com find jbert You onl...

Страница 2: ...I pollution degree 2 environment It is designed to operate within a temperature range of 5 40 C 40 105 F at a maximum relative humidity of 95 and at altitudes of up to 2000 meters Refer to the specifi...

Страница 3: ...the installation media All data and clock input and output connectors of the PG that are not used in your test setup must be terminated with 50 Ohm The instrument can operate from any single phase AC...

Страница 4: ...ct the Agilent IO Control icon in the Windows task bar 5 If you see this icon the revision of the Agilent IO Libraries Suite is 15 5 or later You may wish to connect a keyboard and mouse to your Seria...

Страница 5: ...he rear USB ports of your N4903B 3 When you connect the N4916B for the first time the Windows Found New Hardware Wizard appears and asks whether it should connect to Windows Update Answer No 4 Click N...

Страница 6: ...e alias name to DATA_OUT or AUXDATA_OUT depending on the channel Data Out and Aux Data Out used This is the alias used by the software 3 After clicking OK the Agilent Connection Expert shows a window...

Страница 7: ...these functions 5 Enabling Clock Multiplier Select the Clock Multiplier function to open the Clock Multiplier Connection dialog 7 If you select the Clock Multiplier function a new menu entry is added...

Страница 8: ...For details see the Help and the User Guide Choose a suitable pattern e g PRBS and start the pattern generator 9 After some adjustments of the scope you will see a display like the following 10 In ca...

Страница 9: ...of problems please refer to the section Problems with the N4916B of the Help and the User Guide of your N4903B Preset the instrument state 1 Preset the instrument state 1 Enable Clock Multiplier from...