Error Messages
If the GC displays an LTM Series II system fault, check the appropriate error message below.
Also, check for more details in the GC touchscreen or in the Browser Interface.
•
Select
Diagnostics
on the GC touchscreen.
•
View the configured electronics (controller) board information. See
Diagnostic Information at the GC”
.
•
View the column module information. See
“Accessing LTM Diagnostic Information at the
.
CAUTION
The LTM diagnostics displays include some parameters which should not be altered or
changed. Do not alter or change any settings unless specifically directed by these
instructions. Changing the PID values, for example, can prevent proper operation or damage
the column module.
After reinstalling or replacing a cable connection to an LTM electronics board, power cycle the
GC. The GC checks for proper connections, power usages, and so forth only during power up.
The GC checks for out of range temperature conditions, sudden communications loss, and
similar events during operation.
Table 6
LTM Warning messages
Message
Cause
Effect
LTM Column 1 Changed
A new LTM column was installed.
The column is configured and in use.
LTM Column 1 New
A new LTM column was installed.
The column needs to be configured.
LTM Column 1 Removed
A LTM column is no longer detected.
The LTM column cannot be operated.
LTM Column 2 Changed
A new LTM column was installed.
The column is configured and in use.
LTM Column 2 New
A new LTM column was installed.
The column needs to be configured.
LTM Column 2 Removed
A LTM column is no longer detected.
The LTM column cannot be operated.
LTM Column 3 Changed
A new LTM column was installed.
The column is configured and in use.
LTM Column 3 New
A new LTM column was installed.
The column needs to be configured.
LTM Column 3 Removed
A LTM column is no longer detected.
The LTM column cannot be operated.
LTM Column 4 Changed
A new LTM column was installed.
The column is configured and in use.
LTM Column 4 New
A new LTM column was installed.
The column needs to be configured.
LTM Column 4 Removed
A LTM column is no longer detected.
The LTM column cannot be operated.
LTM Controller 1 Module Communication Failure
Unable to communicate setpoints and
status from the EPC module
Flows will not be controlled to the desired
setpoints.