3
Column Modules
56
User Guide
To Remove a Column Module Assembly from a
Transfer Line Module
This procedure assumes that the module assembly has been removed from the LTM over
door and is cool.
To separate an LTM column module from the transfer line assembly when using a CFT union:
1
Gather the following:
•
Column cutter, wafer (5181-8836, 4/pk)
•
Magnifying loupe, 20X (430-1020)
•
Gloves, heat-resistant (for handling hot parts)
•
Gloves, lint free (to prevent contamination of the column, ferrules, and so forth with skin
oil and dirt)
•
UltiMetal ferrule, appropriate for the column size (see
•
Two 1/4-inch open-end wrenches
•
T20 Torx driver
•
Short metric ruler
•
One 7/16-inch open end wrench
2
Using two wrenches, loosen the column’s internal nut at the CFT union and remove the
fitting from the union. Repeat for the other column connection.
3
Slide the internal nut away from the UltiMetal ferrule, then use a column cutting wafer to
cut the column immediately behind the ferrule. Remove the internal nut from the column
and save for future use. Repeat for the other column connection.
Содержание LTM II Series
Страница 1: ...Agilent LTM Series II Rapid Heating Cooling System for Agilent 8890 GCs User Guide ...
Страница 6: ...6 User Guide ...
Страница 14: ...1 Introduction 14 User Guide ...
Страница 64: ...4 Operation 64 User Guide ...
Страница 72: ...5 Troubleshooting 72 User Guide ...