3
Column Modules
User Guide
55
11
If installing another LTM column module, do so now. If you will not use the open slot(s),
restore the GC to operation as follows:
a
Remove or reconfigure the GC columns as needed.
b
Fill the open slots in the LTM oven door with insulation. (See
.)
c
Using a T-10 Torx driver, loosen the fan bracket mounting screws and remove the fan
bracket. Disconnect the fan from the electronics board.
d
Reinstall the slot cover(s).
e
Close the LTM electronics enclosure.
12
Exit maintenance mode. Select
Finished.
The CFT fittings on the In and Out segment columns can be installed into another CFT fitting
as-is.
If storing the column module, protect the column ends and CFT fittings.