3
Column Modules
User Guide
51
Column module installation is complete.
Normally, leave the GC powered off until after installing the In and Out segments. Leaving the
GC powered off prevents accidental shutdown conditions and heating of some components
while working.
When finished, turn on the GC. The GC will read primary column configuration information
from the column module as if it were a column Smart ID key. However, you will need to
configure other column information, such as the In and Out segment dimensions. See
If the GC detects a problem during startup, turn off the GC. Check the column module and
transfer line heater connections at the electronics board.
LTM top cover
Screws