Maintenance
Putting the Instrument into Storage
Agilent InfinityLab ELSD User Manual
92
Putting the Instrument into Storage
If the instrument is to be stored or not used for an extended period, it is
recommended to follow this procedure:
1
Flush the detector with a mixture of Acetone/Water (50/50) at 1 mL/min for
15 min.
2
Allow the instrument to cool to ambient temperature in STANDBY mode with
the gas supply still connected.
3
Tip the instrument forwards to try and empty the solvent within the
nebulization chamber through the front waste tube (i.e. into the bottle).
4
Pour 10 – 20 mL of acetone into the rear exhaust tube to flush out the internal
solvent trap, collecting any overflow of acetone at the front solvent pipe.
5
Repeat step 3 to drain the acetone.
6
Disconnect the waste bottle.
7
Using the gas supply, blow nitrogen gas through the exhaust to evaporate any
remaining acetone in the solvent trap. Cover the waste tube with tissue paper
to collect any acetone residue.
8
Plug the exhaust, waste tubes and solvent inlet with the plastic caps provided.