Error Information
Module Specific Error Messages
Agilent InfinityLab ELSD User Manual
80
Invalid Nebulizer temperature
Error ID: 22
Invalid nebulizer temperature reading.
Invalid Evaporator temperature
Error ID: 23
Invalid evaporator temperature reading.
Fan failed on cooled evaporator
Error ID: 24
Fans on cooled evaporator module have failed.
Probable cause
Suggested actions
1
Nebulizer heater not connected to the
mainboard.
Please contact your Agilent service representative.
2
Defective nebulizer heater.
Please contact your Agilent service representative.
Probable cause
Suggested actions
1
Evaporator heater not connected to the
mainboard.
Please contact your Agilent service representative.
2
Defective evaporator heater.
Please contact your Agilent service representative.
3
RTD Block to TEC board for cooled ELSD
not connected to the mainboard.
Ensure that the evaporator heater is connected
correctly.
Probable cause
Suggested actions
1
Fan not connected to Peltier assembly.
Please contact your Agilent service representative.
2
Defective Peltier fan.
Please contact your Agilent service representative.