6
Troubleshooting and Diagnostics
General Problems
Agilent InfinityLab ELSD User Manual
72
Display not on, but power connected
Cooled Evaporator Temperature reads zero at
start-up and cannot be changed
Vapor sensor error occurs, but there is no solvent or
vapor leak inside unit
Probable cause
Suggested actions
1
Instrument Power Supply
Refer to local distributor or Agilent Technologies.
2
Faulty display
Refer to local distributor or Agilent Technologies.
Probable cause
Suggested actions
1
Peltier cooler has not initiated correctly
•
Switch detector off then on at the power
socket.
•
Consult Agilent Technologies or your local
agent if problem persists.
Probable cause
Suggested actions
1
Solvent vapor near the front of unit is being
drawn into the unit
Remove any solvent bottle or solvent leak that is
directly in front of the detector.
2
Faulty Vapor sensor
Check the rear vapor sensor is not
damaged/bent.