3000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Segmented Memory optimizes oscilloscope acquisition memory by only
capturing important segments of a waveform based on the trigger
condition and timebase setting. In this example, we have selectively
captured approximately 10 seconds of total acquisition time as shown in
. Note that we can also view the segmented decoded CAN serial
data in the “lister” format, and we can also perform Search & Navigation
operations on the captured segments.
39
Press the
[Serial]
front panel key; then press the
Lister
softkey.
Because we performed the Segmented Memory acquisition based on
triggering on any “form” or flagged “error frame” condition (All Errors),
our Lister display now shows that every captured frame contains some
type of error as shown in
.
Figure 72 Lister display shows errors in every frame from a Segmented Memory
acquisition based on triggering on “All Errors”.