82
3000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Your scope should now be triggering on MOSI data = 03
HEX
with this byte
shown at center- screen similar to
. Let’s now capture a very long
stream of SPI traffic and then manually search through the captured and
decoded record.
30
Press the
[Serial]
front panel key.
31
Press the
Lister
softkey; then turn the Entry knob to select
Serial 1 (SPI)
.
32
Set the scope’s timebase to
20.00 ms/div
.
33
Press
[Run/Stop]
to stop repetitive acquisitions.
34
Press the
Scroll Lister
softkey; then turn the Entry knob to manually
scroll through the lister table.
As you scroll through the data, note that waveforms “track”. This means
that the frame that the arrow points to in the lister table corresponds to
the waveforms that are positioned at center- screen. If you want to zoom
in on a particular SPI frame of data, either push the Entry knob, or press
the
Zoom to Selection
softkey. Let’s now perform an automatic search to find
every occurrence of MOSI data = 03
HEX
. We will then automatically
navigate to each of these occurrences.
35
Push the horizontal position/delay knob to re- position the trigger point
back to center- screen.
36
Set the scope’s timebase to
20.00 ms/div
.
37
Press the
[Search]
front panel key.
38
Press the
Search Edge
softkey; then select
Serial 1 (SPI)
.
Figure 57 Triggering on MOSI data = 03
HEX
.