Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Installation Instructions
7
Figure 6
Aux zone/valve box cable and solenoid valve bracket install
4
Align the solenoid valve bracket so the cables are coming from the bottom
of the bracket and the screws are located at the top over the screw holes in
the frame. See
Figure 6
.
5
Tighten the captive screws into the frame.
6
Connect the MSD heated zone cable (long) to the bracket in the first (A1) or
last (A2) position. Record its position for use during configuration.
7
Route the MSD heated zone cable (long) across the top of the GC oven. Pass
it under any detector gas lines and through the channels to the right side of
the instrument. See
Figure 7
.
Figure 7
Route the MSD heater cable across the GC oven top (view from front of GC)
8
On the left side of the oven top, unfasten the grommet and include the MSD
heated zone long cable. See
Figure 8
. Refasten the grommet.
AI
V1
V2
V3
V4
A2
Solenoid valve
bracket
Bracket captive
screws
MSD heated zone
cable (installed in
first position)
Aux zone/valve box
cable connection at P22