24
Installation Instructions
Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Restore the GC to operating condition
1
Install the top back cover, GC right side cover, electronics cover, pneumatics
cover, and detector cover.
2
Clean up any residual insulation in the oven or third detector box frame
assembly.
3
Plug in the GC and turn on the power.
4
Unlock the GC configuration. Press
[Options]
, select
Keyboard & Display
, and
press
[Enter]
. Scroll to
Hard Configuration Lock
and press
[Off/No]
.
5
On the GC keypad, press
[Configure]
then
[MS/Aux Det]
.
6
Select the new Aux Detector and press [
Enter
].
7
On the
Unconfigured
parameter, press
[Mode/Type]
.
8
Scroll to select the
Aux detector
and press [
Enter
]. Select
Aux 1
if you attached
the MSD heated zone cable to plug A1, or select
Aux 2
if you used plug A2.
9
Press
[Enter]
. A caution message will appear instructing you to reboot.
10
Reboot the GC.
a
Press
[Options]
.
b
Scroll to
Communications
and press
[Enter]
.
c
Scroll to
Reboot the GC?
and press
[On/Yes]
twice to reboot the GC and
have the changes take effect.
11
Configure the detector gas type. Press [
Config
] [
MS/Aux Det
], select the
detector, then press [
Enter
]. Scroll to
Makeup gas type
. If the gas type is
incorrect, press [
Mode/Type
] to change it.
12
Zero the pressure sensors in the EPC module.
a
On the GC keypad, press
[Options]
.
b
Scroll to
Calibration
and press
[Enter]
.
c
Scroll to select the new aux detector and press
[Enter].
d
Scroll to
Zero
line and press [
Info
]. The GC will remind you of the
conditions necessary for zeroing the sensor.
e
Press [
On/Yes
] to zero or [
Clear
] to cancel.