Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Installation Instructions
15
3
Install the
μ
ECD module on the side of the GC by tightening the four captive
screws using a T20 screw driver and finish tightening the screws with the
T20 torque driver.
Figure 19
.
Figure 19
Install the 4 captive screws
4
Install the detector PCA frame and interface board assembly into the third
detector box assembly frame using the two captive screws. See
Figure 20
3rd detector box assembly
3rd µECD installed
Screws