Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Installation Instructions
13
Figure 16
Install the third detector heater cable into the frame
Install the detector
1
Install the interface board onto the detector PCA frame assembly. Slide the
board along the frame so that all hooks in the frame engage the slots in the
board, then secure the board to the frame using the thumbscrew.
Note that there are three sets of hooks on each side of the board. All hooks
must be engaged for the board to be installed securely. See
.
Figure 17
Hooks and thumbscrew
Hooks
Tighten thumbscrew