10
Installation Instructions
Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Prepare the GC left side
1
Use a flat-head screwdriver to remove the sheet metal knockout on the left
side of the GC. See
Figure 11
.
Figure 11
Remove sheetmetal knockout
2
Remove the circular insulation plug and any excess insulation. Use a
flat-bladed screwdriver to create a 1/4-inch hole through the oven
insulation and into the oven. See
Figure 12
.
Figure 12
Prepare mounting hole for the µECD
3
Install the 3rd detector adaptor plate to the side of the GC using four
12 mm M4 screws. See
Hole for 3rd µECD installation
Insulation removed