22
Installation Instructions
Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Reinstall the cover, and install the label and nameplate
1
Slide the 3rd detector box cover over the 3rd detector box and secure with
M4 flathead screws using the T20 torque driver. See
.
2
Place the
μ
ECD caution label on the 3rd detector box cover. See
.
Figure 30
Slide the cover back on and attach the µECD caution label
3
Place an
μ
ECD nameplate on the back of the 3rd detector box assembly next
to the vent tube as shown in
Figure 31
.
Figure 31
µECD nameplate
µECD caution label
µECD nameplate