Configuration and Operation of the Fraction Collector
Optimizing fraction collection
Optimizing fraction collection
Time-based
fraction
collection
•
Time slices
must have a length of at least 0.05 min.
•
Set
# of Fractions
such that length of resulting fractions is at least 0.05 min.
Peak-based
fraction
collection
•
Set threshold and slope values such that length of fractions is at least 0.05
min.
•
Unresolved peaks can be separated using appropriate threshold and slope
values. If two unresolved peaks are to be collected as one fraction, collect
based on threshold only.
•
If the baseline of the chromatogram is below or above 0 mAU, this offset is
not accounted for when triggering peaks using a threshold value. The
threshold value is always added to 0 mAU.