Agilent 1100 Series Fraction Collectors User’s Guide
73
Repairing the Fraction Collector
3
Simple Repairs
8 Connect the finger tight fittings of the inlet /
waste tubing assembly to the ports of the
diverter valve.
IMPORTANT: The tubings must not be bent
up- or downwards. The cables must not be
twisted.
(View from the bottom)
9 A color coded ring on one of the tubings and
the valve body indicates, which cable belongs
to which port.
IMPORTANT: It is absolutely vital to
connect these tubings as described, in
order to maximize their lifetime and
operating security.
10
IMPORTANT: The cables must run into the
ports of the diverter valve in lines parallel
to the horizon.
On completion of this procedure:
•
Re-install the tray(s) in the tray base.
•
Start the instrument.
•
Close the front cover.