Agilent 1100 Series Fraction Collectors User’s Guide
3
Repairing the Fraction Collector
Simple Repairs
Exchanging the Internal Tray
Frequency
When defective
Tools required
None
Parts required
Internal tray
analytical scale
, (G1364-63113)
Internal tray
preparative scale
,(G1364-63114)
Before beginning this procedure:
•
Position the transport unit of the fraction
collector in the “Home Position” (see
•
Remove all installed trays from the tray
base.
•
Turn off the instrument.
1 Locate the internal tray assembly with the
rinse funnel and flow delay sensor in the
bottom of the right front corner of the
instrument.