40
FieldFox User’s Guide
Then
NA
.
About S-parameters
S-parameters (scattering parameters) are used to describe the way a device
modifies a signal. The FieldFox can measure
four S-parameters.
The syntax for
each parameter is described by the following:
S (out | in)
out
= FieldFox receiver port
in
= FieldFox source port
The FieldFox automatically switches the internal source and receivers to make
both forward and reverse measurements. Therefore, the FieldFox can measure
all four S-parameters with a single connection.
When the source comes from port 1, the measurement is said to be in the
forward
direction.
When the source comes from port 2, the measurement is said to be in the
reverse
direction.
S11 and S22 reflection measurements are used to measure the amount of
reflections off the corresponding DUT port. Low reflections means there is a
good impedance match between the source and DUT.
S21 and S12 transmission measurements are used to measure the loss or gain
through a DUT over a specified frequency range. Both ends of the DUT must be
connected to the FieldFox. The FieldFox signal source is transmitted out one of
the test port connectors, through the DUT, and into the other test port connector.
How to measure S-parameters
1. Press
Preset
then
Preset
2. Press
Measure 1
then choose from the following:
o
S11
Reflection measurement at port 1.
o
S21
Forward 2-port transmission measurement.
o
S12
Reverse 2-port transmission measurement. May require an option.
o
S22
Reflection measurement at port 2. May require an option.
OR
select a multi-trace configuration. Learn more on page 42.
3. Press
Freq/Dist
then either
Start
and
Stop
or
Center
and
Span
to enter a
Frequency Range for the measurement.