Data Analysis Features
189
Use
Data - Memory
to show the relative differences between two signals. Use for
comparison of very small signals, such as the S11 match of two connectors.
In the following table, Data/Mem shows both pairs of connectors to have the
same 2 dB difference. However, the second pair of connectors have much better
S11 performance (–50 and –52) and the relative significance is shown in the
Data-Mem values.
Values to compare
Data/Mem
Data-Mem
–10 dB and –12 dB
2 dB
–24 dB
–50 dB and –52 dB
2 dB
–64 dB