![Agilent Technologies FieldFox N9912A Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/agilent-technologies/fieldfox-n9912a/fieldfox-n9912a_user-manual_2868113099.webp)
SA (Spectrum Analyzer) Mode
99
Enter a value between 100 Hz and 3 GHz using the numeric keypad,
▲|▼
arrows, or the rotary knob.
Press
Enter
Other Channel Power Settings
All relevant FieldFox settings are made automatically to ensure the highest
accuracy, such as ResBW, VideoBW, and sweep (SwpAcquisition) speed. These,
and all other SA Mode settings, can be changed manually in a Channel Power
measurement.
Learn about Measurement Preset on page 97.
Learn about Averaging on page 97.
Learn about RRC Weighting on page 103.
Occupied Bandwidth
Occupied Bandwidth measures the power of the displayed frequency span and
displays vertical posts at the frequencies between which the specified percentage
of the power is contained. The frequency span between the two vertical posts is
the Occupied Bandwidth. The Occupied Power, the power that is contained
between the two posts, is also displayed in dBm.
OBW measurement; Chan 190 Downlink; GSM850 Radio Standard
How to select Occupied Bandwidth
Press
Measure__1
Then
Channel Measurements
Then
Occupied BW
When Occupied Bandwidth is selected, the following settings are maintained
from a previous measurement: Center Frequency, Preamp ON|OFF, and RF
Attenuation.
Содержание FieldFox N9912A
Страница 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Страница 134: ...134 N9912A FieldFox User s Guide RF Out and RF In test port cables attached to form a Thru connection ...
Страница 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Страница 195: ...Safety Considerations 195 Inspired Energy Battery ...
Страница 196: ...196 N9912A FieldFox User s Guide ...
Страница 197: ...Safety Considerations 197 ...
Страница 198: ...198 N9912A FieldFox User s Guide ...