SA (Spectrum Analyzer) Mode
89
Then
AutoTrig Time [current setting]
Enter a Auto Trig Time using the numeric keypad.
Then select a multiplier key. Learn about multiplier abbreviations on page 19.
Single / Continuous / Restart
NOTE
The following behavior is unique to SA Mode.
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
or
Run / Hold +/-
button is pressed. Use
Hold /
Single
or to conserve battery power or to allow you to save or analyze a specific
trace.
How to set Single, Continuous, or Restart
Press
Sweep 3
.
Then choose one of the following:
o
Continuous
o
ON
Causes the SA to make continuous sweeps. This is the typical setting
for making measurements. Each
Run / Hold +/-
press causes the SA to
toggle between continuous sweep and Hold.
o
OFF
Causes the SA to Hold. Each
Run / Hold +/-
or
Single
press causes
the SA to make ONE sweep, then Hold. The
Hold
annotation changes to
an arrow
-->
while the sweep occurs.
o
Single
Automatically sets Continuous OFF and causes the SA to make ONE
sweep, then Hold.
o
Restart
Resets the Average counter to 1. When in Hold, causes the SA to
make ONE sweep, then Hold. The Average counter has is used with MaxHold
and MinHold Trace States. Learn more on page 91.
Points
Points is the number of measured frequencies the X-axis. The higher number of
data points, the better the ability to resolve closely spaced signals and the slower
the sweep speed.
How to set number of points
Press
Sweep 3
Then
Points [current setting]
Then choose from:
101 | 201 | 401 | 601 | 801 | 1001
Trace Display States
In SA Mode you can display up to four of the following types of trace states. All
SA settings are applied to all displayed traces.
A color-coded legend for displayed traces is visible in the left pane of the SA
Mode screen:
W
= Clear/Write;
M
= MaxHold;
m
= MinHold;
A
= Average;
V
= View
Содержание FieldFox N9912A
Страница 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Страница 134: ...134 N9912A FieldFox User s Guide RF Out and RF In test port cables attached to form a Thru connection ...
Страница 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Страница 195: ...Safety Considerations 195 Inspired Energy Battery ...
Страница 196: ...196 N9912A FieldFox User s Guide ...
Страница 197: ...Safety Considerations 197 ...
Страница 198: ...198 N9912A FieldFox User s Guide ...