DTF (Distance to Fault) Measurements
35
The following is an example showing how DTF cable correction works:
The DUT is a 100 meter transmission cable. The Cable Loss value is .1
dB/meter. This means that a signal traveling ONE WAY through the cable will
lose 10 dB of power (100 m * .1dB/m). Because the FieldFox performs this
measurement with 1 port, the test signal travels down the cable and then back,
for a total loss of 20 dB.
After a calibration has been performed, for the purpose of illustrating this point,
connect an OPEN to the end of the cable – a maximum-sized fault - for 100%
reflection of the 300 MHz test signal.
Without compensation
for the loss of the cable, a –20 dB response would be
visible at 100 meters, which is the OPEN at the end of the DUT. This is from 10
dB of loss through the cable in each direction.
With compensation
for the loss using the manufacturer’s specification, the
FieldFox compensates the trace as though the signal traveling through 100
meters was increased by +20 dB. Therefore the response will show 0 dB for 100%
reflection.
How to enter Cable Loss and Velocity Factor
Cable Loss and Velocity factor can be entered using one of the following
methods:
Manually enter cable loss and velocity factor for the measurement.
Select or create a cable file which contains the cable loss and velocity factor.
With a DTF measurement present:
Press
Meas Setup 4
.
Then
DTF Cable Specifications
Select
Cable Corr
o
Auto
Use Cable Loss and Velocity Factor values from a Cable file. See “How
to Edit, Save, and Recall a Cable File” below. This will overwrite a manually-
entered value.
o
Man
(Default setting) Manually enter a value for Cable Loss and Velocity
Factor.
Then:
o
Velocity Factor
Using the numeric keypad, enter a value between 0.01 and
1. Then press
Enter
.
o
Cable Loss
Using the numeric keypad, enter a positive Cable Loss value in
dB/m, then press
Enter
.
How to Edit a Cable File
The FieldFox includes many predefined cable files with the manufacturer’s
specifications. You can edit these files or create new cable files using the
following procedure or using the FieldFox Data Link Software. Learn more at:
www.agilent.com/find/fieldsfoxsupport
The Cable correction data survives a Mode Preset and Preset.
With a DTF measurement present:
Содержание FieldFox N9912A
Страница 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Страница 134: ...134 N9912A FieldFox User s Guide RF Out and RF In test port cables attached to form a Thru connection ...
Страница 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Страница 195: ...Safety Considerations 195 Inspired Energy Battery ...
Страница 196: ...196 N9912A FieldFox User s Guide ...
Страница 197: ...Safety Considerations 197 ...
Страница 198: ...198 N9912A FieldFox User s Guide ...