44
N9912A FieldFox User’s Guide
NOTE
Phase is unwrapped by comparing the phase from one data point to the next. If
the phase difference between two data points is greater than 180 degrees, or if
the phase of the first data point is greater than 180 degrees from DC, than the
phase measurement is probably NOT accurate. Increase the resolution setting to
ensure that the phase measurement is accurate.
In Smith and Polar formats, marker format can be changed to display (Mag &
Phase) or (R + jX). Learn how to set marker format on page 139.
Frequency Range
Set the range of frequencies over which you would like to make measurements.
When the frequency range is changed after a calibration is performed, the cal
becomes interpolated. Learn more on page 65.
How to set Frequency Range
This can be done in two ways:
Press
Freq/Dist
Then choose from the following:
1.
Start
and
Stop
frequencies – Specify the beginning and end of the sweep.
2.
Center
and
Span
frequencies - Specify the center frequency and span of
frequencies (half on either side of center).
Follow each by entering a value using the numeric keypad, the
▲|▼
arrows, or
the rotary knob.
After using the
▲|▼
arrows or the rotary knob, press
Enter
. The increment
setting of the arrows is based on the current span and can NOT be changed in
NA Mode.
After using the keypad, select a multiplier key. Learn about multiplier
abbreviations on page 19.
Scale Settings
Adjust the Y-axis scale to see the relevant portions of the data trace. The Y-axis is
divided into 10 graticules.
This setting can be changed at any time without affecting calibration accuracy.
How to set Scale
Press
Scale / Amptd
.
Then choose from the following methods:
1.
Autoscale
Automatically adjusts the Y-axis to comfortably fit the Min and
Max amplitude of the trace on the screen.
2.
Autoscale All
Autoscales all of the traces on the screen, useful only for
multi-trace configurations.
3. Set Scale, Reference Level, and Reference Position
o
Scale
Manually enter a
scale per division
to view specific areas of the trace.
Содержание FieldFox N9912A
Страница 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Страница 134: ...134 N9912A FieldFox User s Guide RF Out and RF In test port cables attached to form a Thru connection ...
Страница 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Страница 195: ...Safety Considerations 195 Inspired Energy Battery ...
Страница 196: ...196 N9912A FieldFox User s Guide ...
Страница 197: ...Safety Considerations 197 ...
Страница 198: ...198 N9912A FieldFox User s Guide ...