SA (Spectrum Analyzer) Mode
75
o
Log
Logarithmic scale (default setting). The Y-axis reference line represents
the specified absolute Reference Level in the current
Unit
selection. Y-axis
graticules show dB below or above the reference line.
o
Lin
Linear scale. The Y-axis reference line represents the specified absolute
Reference Level in the current
Unit
selection. Y-axis graticules show percent
of Reference Level (%) below or above the reference line.
Manually set Scale, Reference Level, and Reference Position.
o
Scale
Manually adjust the scale per division to view specific areas of the
current trace. Enter a value using the numeric keypad, the
▲|▼
arrows, or
the rotary knob. Then press
Enter
.
o
Ref Level
Manually set the value of the reference line. Enter a negative
value by pressing
Run/Hold (+/-)
either before or after typing a value. This
can also change the RF Attenuation setting. See
Attenuation Control –
Auto.
o
Ref Position
Manually set the position of the reference line. Values must be
between 0 (TOP line) and 10 (BOTTOM line). Default position is zero (top).
o
Press
More
, then
Autoscale
Automatically adjusts the Y-axis to comfortably
fit the Min and Max amplitude of the trace on the screen.
How to set Y-axis Units
The Unit setting appears for the reference line, marker readouts, and trigger
level. All Unit choices are available in both Log and Linear Scale Types.
The UNIT setting survives Mode Preset, but NOT Preset.
Press
Scale / Amptd
, then
More
, then
Unit
Then choose from the following:
o
dBm
(default setting),
dBmV
(dB milli Volts),
dBuV
(dB microVolts),
W
(watts),
V
(volts),
A
(amps),
dBmA
(dB milliAmps),
dBuA
(dB microAmps).
Antenna correction units are available ONLY by loading or editing an Antenna
file that contains the desired units setting. Learn more on page 78.
External Gain
When using an external amplifier or attenuator, the SA Mode trace amplitude
values can be offset to compensate for the effect of the external device. This
effectively moves the reference plane of the SA measurement port out to just
beyond the external device. For example, when using an external preamp with
gain of +10 dB, enter this value for External Gain,
and the data trace across the
displayed frequency span will be adjusted down
by 10 dB.
When RF Atten is set to Auto, you may see a change in the RF Attenuation value.
This is an attempt to measure the signal at top of screen (the Reference Level)
without overloading the SA first mixer.
How to set External Gain
Press
Scale/Amptd
.
Then
More
Содержание FieldFox N9912A
Страница 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Страница 134: ...134 N9912A FieldFox User s Guide RF Out and RF In test port cables attached to form a Thru connection ...
Страница 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Страница 195: ...Safety Considerations 195 Inspired Energy Battery ...
Страница 196: ...196 N9912A FieldFox User s Guide ...
Страница 197: ...Safety Considerations 197 ...
Страница 198: ...198 N9912A FieldFox User s Guide ...