18
User’s and Service Guide
E8257D Option HAR
Service
Making the 100 MHz Measurement
Step 1.
In the Measure drop-down menu verify that the Pause at Connect Diagram is
checked. If not, select it.
Step 2.
In the
Measure
drop-down menu select
New Measurement
.
Step 3.
Select
Yes
in the
Do you want to perform a new Calibration and Measurement?
dialog box. The
Agilent E5500 Instrument Connections
window will appear.
Step 4.
Connect the 100 MHz +8 dBm Output on the 70427A rear panel to the Signal
Input on the 70420A. Refer to
.
Step 5.
Turn On the Divider Mode. Refer to
Step 6.
Select
Continue
and allow Phase Noise Measurement to complete.
Step 7.
Compare the estimated mean value of the trace with the noise specification line
(lower line). If the estimated mean value of the trace falls on or below the
specification line, the LNSG is regarded as having met the noise specification.
Step 8.
Print the E5500 Phase Noise Plot.
Title
100 MHz
Agilent E8257D SNxxxxxxxx
Graph Type
Single- sideband Phase Noise (dBc/Hz)
X Scale Minimum
1 Hz
X Scale Maximum
100E+6
Y Scale Maximum
0
Y Scale Minimum
−
170
Normalize Trace Data to
1 Hz Bandwidth
1.
Delta settings are indicated with a bold font.
Table 15
100 MHz Phase Noise System Setup
Parameter
Setting