12
User’s and Service Guide
E8257D Option HAR
Service
Setting up the E5500 Software
Step 1.
If necessary, install the E5500 revision A.03.03 (or later) software.
Step 2.
Open the E5500 Software User Interface.
Step 3.
Select
System
>
Asset Manager
.
Step 4.
Verify that the Agilent E8257D is listed under Source. If not, double-click Source
and click
Next
.
Step 5.
In the list of
Sources
, select
Agilent E8257D
and click
Next
.
Step 6.
In the
Interface
drop-down menu, select
GPIB
.
Step 7.
In the
Address
field, enter the GPIB address of the E8257D. Example, 19.
Step 8.
Verify that the
Library
field identifies
Agilent Technologies VISA
.
Step 9.
Click
Next
>
Finish
.
Step 10. In the list of
Sources
, select
Agilent E8257D
.
Step 11. Select the icon with the green check-mark. Verify that the connection to the
selected asset has been completed. Press
OK
.
Step 12. Verify that the
Library
field identifies
Agilent Technologies VISA
.
Step 13. Close the
Asset Manager
.
Step 14. Verify that the Phase Noise System is connected as shown in
. The
connections from the 70427A should not be made at this time. The user will be
instructed to make the appropriate connections as part of the following
Performance Test procedure.