User’s and Service Guide
7
E8257D Option HAR
Typical Performance
Amplitude Modulation
AM modulation is not recommended in divider mode. In divider mode, ALC Bandwidth is
limited to 100 Hz and dynamic range is restricted below 500 MHz.
Pulse Modulation
Divider Mode On:
Standard Pulse (UNU) can not be used in divider mode. Option UNW can be used.
Divider Mode Off:
Pulse On/Off ratio is degraded by 10 dB from 2.0 to 3.2 GHz in Standard Pulse (UNU).
Option UNW is not affected.
ALC Performance
The ALC bandwidth is limited to 100 Hz in divider mode. Operation with ALC off is not
recommended.
Ramp Sweep
Ramp Sweep is unavailable with Divider Mode On.