14
User’s and Service Guide
E8257D Option HAR
Service
Table 13
10 MHz Phase Noise System Setup
Parameter
Setting
Type and Range Tab:
Measurement Type
Absolute Phase Noise
(using a phase locked loop)
Start Offset Frequency
1 Hz
Stop Offset Frequency
2 MHz (2E+6)
FFT Averages
10
FFT Quality
High Resolution
Swept Quality
High Resolution
Source Tab:
Carrier Source Connected
Test Set
Carrier Source Frequency
10 MHz (10E+6)
Carrier Source Power
+15 dBm
Detector Input Frequency
No entry needed
Reference Source Power
16 dBm
VCO Tune Constant
0.1 ppm x Carrier
0.1 ppm x
10 MHz = 1 Hz/Volt
VCO Tune Range
±5 Volts
VCO Input Resistance
600 Ohms
Cal Tab:
Measure Phase Detector Constant
Measure Phase Detector Constant
Calculate VCO Tune Constant
Measure VCO Tune Constant
Verify calculated phase lock loop
suppression
Check
If Limit is exceeded
Use Theoretical Value
Block Diagram Tab:
Carrier Source
Manual
Downconverter
None
Reference Source
Agilent E8257D
Phase Detector Mode
Automatic Detector Selection
Test Set Tune Voltage Destination
Reference Source
VCO Tune Mode
DC FM
Test Set Tab:
Input Attenuation
0 dB
LNA Low Pass Filter
Auto
LNA Gain
Auto Gain
PLL Integrator Attenuation
0 dB
Downconverter Tab:
Input Attenuation
Auto
IF Gain
Auto
Microwave/Millimeter Band
Microwave (0 to 26.5 GHz)
Reference Chain Reference
10 MHz
Graph Tab:
Title
10 MHz Agilent E8257D SNxxxxxxxx