User’s and Service Guide
E8257D Option HAR
Operation
Operation
E8257D Option HAR requires firmware revision
≥
C.04.94.
To turn on the frequency divider circuitry (Divider Mode) select
[Frequency]
>
MORE (1 of 3)
>
MORE (2 of 3)
>
Optimize noise below 500 MHz
> toggle
On
or
Off
.
The normal preset state is Off.
The SCPI command is as follows:
[:SOURce]: LFDivider ON|OFF|1|0
[:SOURce]: LFDivider?
Figure 2
Frequency Divider Circuitry
E8257D Option HAR Using the E5500 Phase Noise Measurement
Subsystem
The E5500 software currently does not recognize Option HAR. To improve reference phase
noise when using your E8257D Option HAR with the E5500, manually turn the Divider
Mode On (refer to
), when the “Agilent E5500 Instrument Connections” window
appears.