Status
Register
Structure
Status
byte
totals
three
status
registers
which
indicate
the
internal
condition
of
an
instrument.
Figure
5-3
shows
the
status
reporting
structure
of
the
E5100A/B .
Figure
5-3.
Status
Reporting
Structure
The
E5100A/B
has
a
status
reporting
system
to
report
the
condition
of
the
instrument.
Status
bytes
consists
of
8-bit
registers
,
each
bit
represents
specic
instrument
conditions
.
The
value
of
the
status
byte
can
be
read
by
using
SPOLL(717)
statement
from
an
external
controller
.
This
command
reads
value
directly
from
the
E5100A/B
without
being
set
to
remote
.
So
,
you
can
operate
front
panel
keys
while
a
controller
is
reading
the
status
byte
.
Contents
of
the
status
byte
can
also
be
read
by
using
the
*STB?
command.
Reading
the
status
byte
has
no
eect
on
the
contents
of
the
status
byte
.
T
able
5-1
shows
contents
of
status
byte
.
5-4
Using
Status
Reporting
Function
Содержание E5100A
Страница 1: ...Agilent E5100A B Network Analyzer Programming Manual ...
Страница 2: ......
Страница 5: ...c Copyright 1995 1996 1997 1998 2000 2001 Agilent Technologies Japan Ltd ...
Страница 8: ......
Страница 36: ......
Страница 46: ......
Страница 48: ...Figure 3 1 Simpli ed Data Processing Flow 1 3 2 Data Processing and T ransfer ...
Страница 58: ......
Страница 76: ......
Страница 105: ...610 RETURN 620 630 Ending ENDING PROGRAM 640 ASSIGN Dt TO 650 660 END Application Sample Programs 7 7 ...
Страница 112: ......
Страница 232: ......
Страница 246: ......
Страница 250: ......
Страница 277: ...110 PRINT Pole left Pole1 dB Fp1 Hz 120 PRINT Pole right Pole2 dB Fp2 Hz 130 END Waveform Analysis Commands D 27 ...
Страница 299: ...110 END Waveform Analysis Commands D 49 ...
Страница 302: ......
Страница 322: ......
Страница 350: ......
Страница 352: ......
Страница 362: ......
Страница 368: ......
Страница 372: ......