3.
Denes
the
middle
point
between
f
s1
and
f
s2
to
f
s
.
4.
Searches
for
the
x
2
dB
above
points
on
both
sides
,
and
denes
the
rst
found
left
and
right
side
points
as
f
p1
and
f
p2
,
respectively
.
5.
Denes
the
middle
point
between
f
p1
and
f
p2
as
f
p
.
Note
If
there
is
no
corresponding
peak
in
the
range
,
OUTPRESF?
returns
0
for
all
parameters
.
If
the
maximum
peak
cannot
be
found,
OUTPRESF?
returns
0
for
f
s
,
f
s1
,
and
f
s2
.
If
the
minimum
negative
peak
cannot
be
found,
OUTPRESF?
returns
0
for
f
p
,
f
p1
,
and
f
p2
.
Specify
the
negative
value
for
x
1
and
positive
value
for
x
2
.
Examples
10
ASSIGN
@E5100
TO
717
20
OUTPUT
@E5100;"FMT
LOGM;
CENT
60.06MHz;
SPAN
20kHz"
30
CALL
Sweep(1)
!
Goes
to
sub
routine.
(See
OUTPFILT?)
40
OUTPUT
@E5100;"ANAOCH1;ANARFULL;ANAODATA"
50
OUTPUT
@E5100;"OUTPRESF?
-3dB,3dB"
60
ENTER
@E5100;Fs,Fp,Fs1,Fs2,Fp1,Fp2
70
"Series-Resonant:";Fs;"[Hz]"
80
"Parallel-Resonant:";Fp;"[Hz]"
90
END
Waveform
Analysis
Commands
D-41
Содержание E5100A
Страница 1: ...Agilent E5100A B Network Analyzer Programming Manual ...
Страница 2: ......
Страница 5: ...c Copyright 1995 1996 1997 1998 2000 2001 Agilent Technologies Japan Ltd ...
Страница 8: ......
Страница 36: ......
Страница 46: ......
Страница 48: ...Figure 3 1 Simpli ed Data Processing Flow 1 3 2 Data Processing and T ransfer ...
Страница 58: ......
Страница 76: ......
Страница 105: ...610 RETURN 620 630 Ending ENDING PROGRAM 640 ASSIGN Dt TO 650 660 END Application Sample Programs 7 7 ...
Страница 112: ......
Страница 232: ......
Страница 246: ......
Страница 250: ......
Страница 277: ...110 PRINT Pole left Pole1 dB Fp1 Hz 120 PRINT Pole right Pole2 dB Fp2 Hz 130 END Waveform Analysis Commands D 27 ...
Страница 299: ...110 END Waveform Analysis Commands D 49 ...
Страница 302: ......
Страница 322: ......
Страница 350: ......
Страница 352: ......
Страница 362: ......
Страница 368: ......
Страница 372: ......