POLE?
Returns
the
stimulus
and
value
of
the
rst
negative
peak
found
on
each
side
of
the
maximum
point
that
are
below
the
specied
value
from
the
maximum
peak.
(Query
only)
Syntax
POLE?
D
Where
,
Register
P
arameter
Description
0
D
Dierence
from
the
maximum
peak.
Query
Response
x
1
,
stim
1
,
x
2
,
stim
2
Where
,
Register
P
arameter
Description
0
x
1
Left
negative
peak
value
.
1
stim
1
Stimulus
of
x
1
.
2
x
2
Right
negative
peak
value
.
3
stim
2
Stimulus
of
x
2
.
Figure
D-9.
POLE?
Note
If
the
search
fails
,
the
analyzer
returns
0 .
Give
the
command
parameter
as
a
negative
value
.
F
or
instance
,
to
specify
50
dB
down
from
the
maximum
peak
as
a
reference
level,
the
parameter
is
-50 .
Examples
ASSIGN
@E5100
TO
717
OUTPUT
@E5100;"ANAOCH1;ANARFULL;ANAODATA"
OUTPUT
@E5100;"POLE?
-50"
ENTER
@E5100;X1,S1,X2,S2
"LEFT
:";X1;"[dB]";S1;"[Hz]"
"RIGHT:";X2;"[dB]";S2;"[Hz]"
END
Waveform
Analysis
Commands
D-21
Содержание E5100A
Страница 1: ...Agilent E5100A B Network Analyzer Programming Manual ...
Страница 2: ......
Страница 5: ...c Copyright 1995 1996 1997 1998 2000 2001 Agilent Technologies Japan Ltd ...
Страница 8: ......
Страница 36: ......
Страница 46: ......
Страница 48: ...Figure 3 1 Simpli ed Data Processing Flow 1 3 2 Data Processing and T ransfer ...
Страница 58: ......
Страница 76: ......
Страница 105: ...610 RETURN 620 630 Ending ENDING PROGRAM 640 ASSIGN Dt TO 650 660 END Application Sample Programs 7 7 ...
Страница 112: ......
Страница 232: ......
Страница 246: ......
Страница 250: ......
Страница 277: ...110 PRINT Pole left Pole1 dB Fp1 Hz 120 PRINT Pole right Pole2 dB Fp2 Hz 130 END Waveform Analysis Commands D 27 ...
Страница 299: ...110 END Waveform Analysis Commands D 49 ...
Страница 302: ......
Страница 322: ......
Страница 350: ......
Страница 352: ......
Страница 362: ......
Страница 368: ......
Страница 372: ......