Troubleshooting
Booting Process Troubleshooting
Step 1.
Splash Screen
The splash screen is displayed with Agilent logo as shown in Figure 2-8.
If the splash screen is displayed, you can assume that the A50/A60 CPU Module is
functioning correctly.
NOTE
While the splash screen is displayed, if you want to run the BIOS setup utility, push F2 key
as soon as in the screen. The password to enter BIOS setup unility is agt0nly ( 0 is Zero,
not o).
Figure 2-8
Splash Screen
Step 2.
Boot up Screen
The Boot up screen is displayed as shown in Figure 2-9.
Without choose any choice, the system will continue boot up process after 3 second.
NOTE
While the Boot up screen is displayed, if you want to do system recovery , please select
“Agilent Recovery System” as soon as in the screen.
Figure 2-9
Boot up Screen
Step 3.
Windows boot screens
The Windows boot screens are displayed. The Windows boot screens consists of two
screens. Each screen is displayed in the order as shown in Figure 2-10. If the Windows
Содержание E5061B
Страница 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Страница 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Страница 100: ...Chapter 4 99 Replacement Procedure Front Frame Replacement 4 Replacement Procedure Figure 4 24 Front Frame Removal ...
Страница 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Страница 116: ...115 E Power Requirement D Power Requirement ...
Страница 118: ...Appendix D 117 Power Requirement Preparation for Power Supply E Power Requirement Figure D 1 Power cable options ...
Страница 123: ...122 AppendixE Messages ...