Chapter 2
Troubleshooting
To Troubleshoot the Instrument
To Troubleshoot the Instrument
This section describes basic procedural flow of troubleshooting when servicing the
E5061B. The primary procedural tool in this section is the flowchart. The flowchart
contains entire troubleshooting path from a failure symptom to the isolation of faulty
assembly, and will direct you to the completion of repair in an ordinary manner through the
possible failure symptoms. Reference letters (Yes/No) on the flowcharts point to
procedural steps that briefly explain the troubleshooting method to be performed next.
Primary Trouble Isolation
The primary trouble isolation procedure can be performed without disassembling the
E5061B. Figure 2-2 shows the trouble isolation flow chart.
Step 1.
Turn the instrument power on
About a few minutes after the E5061B is turned on, the measurement view is displayed on
the screen. The display on the screen should be similar to Figure 2-12, “Measurement
view,” on page 29.
Step 2.
Check the display
•
If no display appears on the LCD after the E5061B is turned on, go to “No Display
Troubleshooting” on page 22.
•
If the E5061B stops in booting process despite something being displayed on the LCD,
go to “Booting Process Troubleshooting” on page 25.
•
The power-on self test is performed once automatically after the E5061B measurement
view is displayed. If the power-on self test fails, go to “Troubleshooting Using
Diagnostic Test” on page 30.
Step 3.
Check the basic function
If the front-panel/keyboard/mouse controls, LCD display, data storage, remote interface or
another function (except for measurement part) does not work correctly, go to “Function
Specific Troubleshooting” on page 36.
Step 4.
Check the measurement function
If the instrument fails performance tests, go to “Performance Test failure Troubleshooting”
on page 42.
If the measurement function does not work correctly, perform the diagnostic test provided
in the E5061B's service function. When the diagnostic test fails, go to “Diagnostic Test
Failure Troubleshooting” on page 33.
NOTE
The diagnostic test includes some unique measurement function tests in addition to the
tests that are common to the power-on self test. Thus, it is necessary to perform the
diagnostic test even if the power-on self test passed.
Содержание E5061B
Страница 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Страница 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Страница 100: ...Chapter 4 99 Replacement Procedure Front Frame Replacement 4 Replacement Procedure Figure 4 24 Front Frame Removal ...
Страница 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Страница 116: ...115 E Power Requirement D Power Requirement ...
Страница 118: ...Appendix D 117 Power Requirement Preparation for Power Supply E Power Requirement Figure D 1 Power cable options ...
Страница 123: ...122 AppendixE Messages ...