Chapter 2
23
Troubleshooting
No Display Troubleshooting
2. T
ro
uble
shoo
tin
g
Step 1.
Check fan operation and DC monitor LED
If the rear panel fan (blower) doesn't run, a failure in power supply is assumed. Remove the
E5061B outer cover and check if the following LEDs light:
•
+3.3 V and +5 V dc monitor LEDs on A50/A60 CPU Module as shown in Figure 2-4
•
LEDs on DSP module as shown in Figure 2-5
Figure 2-4
LEDs on CPU module
Figure 2-5
LEDs on DSP module
NOTE
To check all the outputs of the power supply, measure the dc voltages at the output lead
connectors with a DMM. The dc output voltages and lead color information is provided in
the module cover label of the power supply.
Содержание E5061B
Страница 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Страница 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Страница 100: ...Chapter 4 99 Replacement Procedure Front Frame Replacement 4 Replacement Procedure Figure 4 24 Front Frame Removal ...
Страница 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Страница 116: ...115 E Power Requirement D Power Requirement ...
Страница 118: ...Appendix D 117 Power Requirement Preparation for Power Supply E Power Requirement Figure D 1 Power cable options ...
Страница 123: ...122 AppendixE Messages ...