Chapter 2
19
Troubleshooting
How to exit from the E5061B Measurement View
2. T
ro
uble
shoo
tin
g
How to exit from the E5061B Measurement View
You need to exit from the E5061B Measurement View to perform some troubleshooting.
The following is the procedure to exit from the E5061B Measurement View.
Step 1.
Connect the mouse and external keyboard to the connectors on the E5061B rear panel.
Step 2.
Turn the instrument on.
Step 3.
Press
key.
Step 4.
Click
Service Menu
-
Exit
. Exit dialog box is as shown in Figure 2-1.
Figure 2-1
Exit dialog box
Step 5.
Click
OK
in
Exit
Menu. Then the E5061B exit the Measurement View, then.windows
desktop screen appears.
NOTE
If you wish to return to the Measurement View, double-click "Network Analyzer" icon.
NOTE
If you need to shut down the E5061B and again turn on, perform in accordance with the
following procedure.
a.
To get "Start" menu bar displayed, move the pointer to the bottom of the screen with
mouse.
b.
Click "Start" and "Shut Down..." in the pull down menu. "Shut Down Windows" dialog
box opens.
c.
Select "Shut down" button in the pull down menu.
d.
Click "OK" button in the dialog box.
Содержание E5061B
Страница 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Страница 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Страница 100: ...Chapter 4 99 Replacement Procedure Front Frame Replacement 4 Replacement Procedure Figure 4 24 Front Frame Removal ...
Страница 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Страница 116: ...115 E Power Requirement D Power Requirement ...
Страница 118: ...Appendix D 117 Power Requirement Preparation for Power Supply E Power Requirement Figure D 1 Power cable options ...
Страница 123: ...122 AppendixE Messages ...