342
Chapter 5
Language Reference
SENSe Subsystem
Waveform—Averaging State
[:SENSe]:WAVeform:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:WAVeform:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST:
Off
Remarks:
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Averaging Mode
[:SENSe]:WAVeform:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:WAVeform:AVERage:TCONtrol?
Select the type of termination control used for the averaging function.
This determines the averaging action after the specified number of
‘sweeps’ (average count) is reached.
Exponential - Each successive data acquisition after the average
count is reached, is exponentially weighted and combined with the
existing average.
Repeat - After reaching the average count, the averaging is reset and
a new average is started.
Factory Preset
and *RST:
Exponential
Remarks:
To use this command, the appropriate mode should be
selected with INSTrument:SELect.