Chapter 5
187
Language Reference
CALCulate Subsystem
PDC Mode - <measurement> key words
• ACP - no markers
• EVM - markers available
• OBW - no markers
• SPECtrum - markers available
• WAVeform - markers available
W-CDMA (3GPP) Mode - <measurement> key words
• ACP - no markers
• CDPower - markers available
• CHPower - no markers
• EVMQpsk - markers available
• IM - markers available
• MCPower - no markers
• OBW - no markers
• PSTatistic - markers available
• RHO - markers available
• SEMask - markers available
• SPECtrum - markers available
• WAVeform - markers available
W-CDMA (Trial & Arib) Mode - <measurement> key words
• ACP - no markers
• CDPower - markers available
• CHPower - no markers
• EVMQpsk - markers available
• PSTatistic - markers available
• RHO - markers available
• SPECtrum - markers available
• WAVeform - markers available
Example:
Suppose you are using the Spectrum measurement. To position marker
2 at the maximum peak value of the trace that marker 2 is currently on,
the command is:
:CALCulate:SPECtrum:MARKer2:MAXimum
You must make sure that the measurement is completed before trying
to query the marker value. Use the MEASure or READ command
before using the marker command. This forces the measurement to
complete before allowing the next command to be executed.
Содержание E4406A VSA Series
Страница 4: ...4 ...
Страница 59: ...59 2 Programming Fundamentals ...
Страница 76: ...76 Chapter2 Programming Fundamentals Using the Instrument Status Registers Overall Status Register System ...
Страница 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Страница 125: ...125 3 Programming Examples ...
Страница 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Страница 165: ...165 4 Programming Command Cross References ...
Страница 379: ...379 6 Error Messages ...
Страница 412: ...412 Chapter6 Error Messages Error Message Descriptions ...