Chapter 5
Language Reference
SENSe Subsystem
Power vs. Time—Resolution BW
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution]?
Set the resolution BW. This is an advanced control that normally does
not need to be changed. Setting this to a value other than the factory
default, may cause invalid measurement results.
Factory Preset
and *RST:
500 kHz
Range:
1 kHz to 5 MHz
Default Unit:
Hz
Remarks:
You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—RBW Filter Type
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution]:TYPE
FLATtop|GAUSsian
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution]:TYPE?
Select the type of resolution BW filter. This is an advanced control that
normally does not need to be changed. Setting this to a value other than
the factory default, may cause invalid measurement results.
Flattop - a filter with a flat amplitude response, which provides the
best amplitude accuracy.
Gaussian - a filter with Gaussian characteristics, which provides the
best pulse response.
Factory Preset
and *RST:
Gaussian
Remarks:
You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Содержание E4406A VSA Series
Страница 4: ...4 ...
Страница 59: ...59 2 Programming Fundamentals ...
Страница 76: ...76 Chapter2 Programming Fundamentals Using the Instrument Status Registers Overall Status Register System ...
Страница 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Страница 125: ...125 3 Programming Examples ...
Страница 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Страница 165: ...165 4 Programming Command Cross References ...
Страница 379: ...379 6 Error Messages ...
Страница 412: ...412 Chapter6 Error Messages Error Message Descriptions ...