
Chapter 5
203
Language Reference
CALibration Subsystem
Calibrate the Nominal System Gain
:CALibration:GAIN:CSYStem
:CALibration:GAIN:CSYStem?
Calculate the current system gain correction for nominal settings. That
is, with 10 dB attenuation, 500 MHz center frequency, 0 dB IF gain and
the prefilter off.
Front Panel
Access:
System, Alignments, Align Subsystem, IF
Align the IF
:CALibration:GIF
:CALibration:GIF?
Performs the IF group of alignments. The query returns a 0 if the
alignments occurred without problems.
Front Panel
Access:
System, Alignments, Align Subsystem, Align IF
Align the RF
:CALibration:GRF
:CALibration:GRF?
Performs the RF group of alignments. The query returns a 0 if the
alignments occurred without problems.
Front Panel
Access:
System, Alignments, Align Subsystem, Align RF
Align the Image Filter Circuitry
:CALibration:IMAGefilter
:CALibration:IMAGefilter?
Align the eight image filter tuning DACs.
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Diagnostics
Содержание E4406A VSA Series
Страница 4: ...4 ...
Страница 59: ...59 2 Programming Fundamentals ...
Страница 76: ...76 Chapter2 Programming Fundamentals Using the Instrument Status Registers Overall Status Register System ...
Страница 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Страница 125: ...125 3 Programming Examples ...
Страница 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Страница 165: ...165 4 Programming Command Cross References ...
Страница 379: ...379 6 Error Messages ...
Страница 412: ...412 Chapter6 Error Messages Error Message Descriptions ...