Chapter 5
299
Language Reference
SENSe Subsystem
Channel Power Measurement
Commands for querying the channel power measurement results and
for setting to the default values are found in the
. The equivalent front panel keys for the
parameters described in the following commands, are found under the
Meas Setup
key, after the
Channel Power
measurement has been selected
from the
MEASURE
key menu. CHPower used instead of the more
std-compliant CPOWer, as that syntax was already used for Carrier
Power measurement (but has since been renamed).
Channel Power—Average Count
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST:
20
Range:
1 to 10,000
Remarks:
You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or Basic mode to use
this command. Use INSTrument:SELect to set the
mode.
Channel Power—Averaging State
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST:
On
Remarks:
You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or Basic mode to use
this command. Use INSTrument:SELect to set the
mode.
Содержание E4406A VSA Series
Страница 4: ...4 ...
Страница 59: ...59 2 Programming Fundamentals ...
Страница 76: ...76 Chapter2 Programming Fundamentals Using the Instrument Status Registers Overall Status Register System ...
Страница 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Страница 125: ...125 3 Programming Examples ...
Страница 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Страница 165: ...165 4 Programming Command Cross References ...
Страница 379: ...379 6 Error Messages ...
Страница 412: ...412 Chapter6 Error Messages Error Message Descriptions ...