3 Common AssayMAP Bravo error messages
Could not dock gripper and other communication errors
8
AssayMAP Bravo Platform Error Recovery Guide
Could not dock gripper and other communication errors
Error message examples
In addition to the errors shown in these examples, some of the other potential
communication and homing errors are:
•
Could not home the X-axis or Y-axis.
•
Could not align X-axis or Y-axis motor.
•
Error in moving to position on X-axis or Y-axis.
•
Error homing the X-axis or Y-axis: Axis timed out during commutation.
•
No response received from the pipette controller.
•
Error retrieving homed status of X-axis or Y-axis while preparing a move.
The exact wording of the error message can vary depending on the Bravo model (for
example, G5563A or G5523A) and the operation that the Bravo device was performing
when the error occurred.
This type of error can occur if you attempt to run an AssayMAP application or utility
after power cycling the Bravo Platform without first exiting the Protein Sample Prep
Workbench and the VWorks software.
A communication can also occur if someone turns off the power to the AssayMAP
Bravo Platform or a power or communication cable is disconnected or loose.
Use the following procedure to resolve any of these types of errors.