3 Common AssayMAP Bravo error messages
Robot disable is active messages
10
AssayMAP Bravo Platform Error Recovery Guide
5
Start the Protein Sample Prep Workbench, and open the
Utility Library
.
6
In the
Utility Library
, locate the
System Startup/Shutdown
utility, and then click
Utility
.
Follow the instructions for running the Startup protocol.
Robot disable is active messages
Error message examples
Note:
The wording in a given
robot disable is active
error message may not match the
examples included here.
A
robot disable is active
error message opens anytime you press the red button on the
emergency-stop pendant or interrupt the Light Curtain, resulting in an emergency stop.
The actual wording in the error message depends on the operation that was in progress
at the time the interlock circuit was tripped.