9
3 Common AssayMAP Bravo error messages
Could not dock gripper and other communication errors
AssayMAP Bravo Platform Error Recovery Guide
Resolving a communication error
IMPORTANT
Make sure that you run the System Startup utility when starting up the AssayMAP
Bravo Platform.
To resolve a communication error:
1
In the
AssayMAP Bravo Error
dialog box, click
Abort.
2
In the
Scheduler Paused
dialog box that opens, click
Abort process
.
3
Exit the Protein Sample Prep Workbench and the VWorks software.
4
Power cycle the device as follows:
a
Turn off the AssayMAP Bravo Platform and accessories.
b
Wait a minute, and then turn on the AssayMAP Bravo Platform and
accessories.