4
Troubleshooting
40
Mass Spec Bench User Guide
Temperature alarm does not turn off.
Open the Mass Spec Bench door. Make sure that
both of the cooling fans are blowing inside the
enclosure. Check the back of the Mass Spec Bench
to make sure that the air intakes for the cooling fans
are not blocked. Make sure that adequate cool air is
available to the Mass Spec Bench. Call your Agilent
service representative if the lab temperature is
within specifications.
Mass Spec Bench poorly muffles the sound
of the pump.
Make sure that the front door is closed tightly and
securely in place.
Green LED does not light.
Check that power is supplied to Mass Spec Bench
and the power switch is on. If it is, call your Agilent
service representative.
Fans do not turn.
Check that power is supplied to Mass Spec Bench
and the power switch is on. If it is, call your Agilent
service representative. Do not continue to use your
rough pump in the Mass Spec Bench.
The temperature alarm does not sound even
though the room temperature is above 35 °C
(95 °F) and the red over-temperature LED is
lit.
Check that power is supplied to Mass Spec Bench
and the power switch is on. If it is, call your Agilent
service representative. Do not continue to use your
rough pump in the Mass Spec Bench— the alarm
may have failed.
The red over-temperature LED is not lit even
though the room temperature is above 35 °C
(95 °F) and the alarm is sounding.
Check that power is supplied to Mass Spec Bench
and the power switch is on. If it is, call your Agilent
service representative. Do not continue to use your
rough pump in the Mass Spec Bench— the LED may
have failed.
The temperature alarm is not sounding and
the red over-temperature LED is not lit even
though the room temperature is above 35 °C
(95 °F).
Check that power is supplied to Mass Spec Bench
and the power switch is on. If it is, call your Agilent
service representative. Do not continue to use your
rough pump in the Mass Spec Bench— the
thermostat may have failed.
Table 4
Problems and solutions (continued)
Problem
Solution
Содержание Agilent G3215A
Страница 1: ...Agilent Technologies Agilent G3215A Mass Spec Bench User Guide Revision A May 2011...
Страница 4: ...4 Mass Spec Bench User Guide...
Страница 25: ...Installation 2 Mass Spec Bench User Guide 25 Figure 15 Single pump orientation Figure 16 Dual pump orientation...
Страница 30: ...2 Installation 30 Mass Spec Bench User Guide Figure 21 Extending the pump tray feet...
Страница 35: ...Operation 3 Mass Spec Bench User Guide 35 Figure 26 Fan power switch...
Страница 41: ...Troubleshooting 4 Mass Spec Bench User Guide 41...