Operation
38
Mass Spec Bench User Guide
provides easier access to the drain port. Be careful not to slide out the
pump too far.
7
Remove the oil drain plug from your rough pump. Refer to your pump
documentation. See
Figure 27
.
8
Reinstall the drain plug when the oil has completely drained.
9
Dispose of the oil according to local safety and environmental guidelines.
10
Return the rough pump(s) to their usual position and remove the fill port
plug.
11
Use a funnel to fill the pump. Refer to the pump documentation for more
information.
12
After the rough pump have been filled, replace the port plug.
13
Close the access door.
C A U T I O N
Wipe up any spilled or excess oil.
WA R N I N G
To avoid creating a fire hazard:
• Never put a paper towel or other absorbent material under the rough pump.
• Always remove from the enclosure any oil soaked or contaminated towels used
for cleaning.