
3
Operation
36
Mass Spec Bench User Guide
To handle temperature alarms
The red LED on the front panel turns on and the alarm sounds when:
•
The air flow to the Mass Spec Bench is restricted.
•
The room temperature is above 35 °C (95 °F).
•
If the red LED is lit:
•
Open the access door to allow air to cool the pump.
•
Check that both fans are turning and that air flow to the Mass Spec
Bench is not restricted.
You cannot turn off the alarm. It stops when the temperature returns to an
acceptable level.
C A U T I O N
The Mass Spec Bench must not recirculate hot air. Always position the back end of the
Mass Spec Bench where it can draw cool air.
Содержание Agilent G3215A
Страница 1: ...Agilent Technologies Agilent G3215A Mass Spec Bench User Guide Revision A May 2011...
Страница 4: ...4 Mass Spec Bench User Guide...
Страница 25: ...Installation 2 Mass Spec Bench User Guide 25 Figure 15 Single pump orientation Figure 16 Dual pump orientation...
Страница 30: ...2 Installation 30 Mass Spec Bench User Guide Figure 21 Extending the pump tray feet...
Страница 35: ...Operation 3 Mass Spec Bench User Guide 35 Figure 26 Fan power switch...
Страница 41: ...Troubleshooting 4 Mass Spec Bench User Guide 41...